AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 1012633-39
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/09/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents from this lot.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The investigation was unable to determine a conclusive cause for the reported difficulties; however the additional patient effect of surgical procedure appears to be related to circumstances of the procedure.
 
Event or Problem Description
It was reported the procedure was to treat a de novo lesion in the right external iliac artery.The lesion was pre-dilated with a 9.0 mm balloon catheter successfully.The 10 x 39 mm omnilink stent delivery system (sds) was advanced and deployed successfully.However, intravascular ultrasound (ivus) noted that the proximal end of the stent implant had folded over.A balloon catheter was advanced in an attempt to mend the stent, but the stent became mangled.A cut down emergency procedure had to be performed to extract the mangled stent from the patient anatomy.The vessel was treated with endarterectomy and completed with a patch.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6125703
Report Number2024168-2016-08321
Device Sequence Number10878906
Product Code NIO
UDI-Device Identifier08717648178597
UDI-Public(01)08717648178597(17)180930(10)5042841
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2018
Device Catalogue Number1012633-39
Device Lot Number5042841
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2016
Initial Report FDA Received Date11/23/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient Weight105
-
-