Model Number S7
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/28/2016
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.Device manufacturing date unavailable.A medtronic representative went to the site to test the equipment.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.
 
Event Description
A site representative reported that during a spinal fusion procedure the representative was unable to connect the navigation system to the imaging system.The site representative attempted to troubleshoot the system by removing the panels on the navigation system to try to bypass the bulkhead without resolution.There was a reported delay to the procedure of less than 1 hour due to this issue.The surgeon opted to complete the procedure without the use of the navigation system.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
amos jarrette
826 coal creek circle
louisville, CO 80027
7208902082
MDR Report Key6126279
MDR Text Key60866701
Report Number1723170-2016-05537
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/12/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-