A portion of the lead assembly including the electrodes was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.What appeared to be white deposits were observed in various locations.Energy dispersion spectroscopy (provides chemical or element identity/composition analysis) was performed on the deposit observed on the outer silicone tubing and identified the deposit as containing silicon, phosphorus, sodium, magnesium and calcium.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portion were performed, during the visual analysis, with no discontinuities identified.Based on the findings in the lab, there is no evidence to suggest discontinuities in the returned portion of the device.
|