It was reported that high lead impedance was observed on patient¿s vns system.Review of manufacturing records confirmed all tests passed for the concerned lead prior to distribution.X-ray was taken and sent to the manufacturer for review.It's unable to verify correct location of the generator due to x-rays partly available.The generator is located on the left side, close to the clavicle.The electrode pin is fully inserted and the feed through wires are intact.There is a suspicion of insulation break at 1 cm from the connector booth.No sharp angles are present no other abnormalities observed on the lead body.The electrodes seem to be aligned with the vagus nerve.No arguments for lead dislocation.Strain relief is present and placed according to the labeling.Two tie-downs are present, but not placed according labeling.Strain relief loop is not present.Based on the images, the cause of the high impedance could not be confirmed, but there is a suspicious area close to the generator.No patient adverse events were reported.No know surgical interventions were performed to date.
|