COOK INC UNKNOWN Back to Search Results
Model Number N/A
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Appropriate Term/Code Not Available (3191)
Patient Problems Perforation of Vessels (2135); No Known Impact Or Consequence To Patient (2692)
Event Date 11/04/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).The event is currently under investigation.
 
Event Description
As reported to customer relations: "an ivc filter retrieval was being performed on a difficult retrieval due to one of the filter legs had penetrated the vena cava wall.The exact identity of the filter and date of implant is unknown.The patient was not symptomatic.((b)(4)) the physician was having to put a lot of pressure on the sheath to try to collapse the filter and during manipulation the hub of the sheath started to separate from the body of the sheath.The retrieval was successful with no harm to the patient.((b)(4))".
 
Manufacturer Narrative
Adverse event/product problem: adverse event (b)(4).No imaging was provided to assist the investigation and it would be inappropriate to speculate at what may or may not have occurred based on the limited information made available to us concerning filter perforation and difficult filter retrieval.Filter implant period is unknown, but patient was not symptomatic.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8128294891
MDR Report Key6132373
MDR Text Key61082900
Report Number1820334-2016-01420
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-