|
This report has been identified as b.Braun inc.Internal report # (b)(4).Three unused sets were received for evaluation.The samples were functionally tested for luer tapers, functional fit, clear passage, and leakage per the specification, all with passing results.There were no leakages observed.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or nonconformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.Based on the results of this investigation, no specific conclusions can be made regarding the cause of the reported event.The returned sets met requirements according to specification, and the reported failure could not be reproduced.If additional pertinent information becomes available a follow-up report will be filed.Note: this case is being filed retrospectively as a result of a review of recent customer complaint information.Based on additional information and details provided in another complaint case, it was determined that this case is reportable in accordance with the requirements of 21 cfr 803.B.Braun has conducted a retrospective review for all complaints of a similar nature in accordance with internal procedure (b)(4).
|