Catalog Number IAP-0435 |
Device Problem
Difficult to Interrogate (1331)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/08/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported via a field service report: (b)(4).Symptom: loss of trigger in autopilot mode; could not "re-acquire." findings / action taken: ran for 1 hour in all trigger in ecg and ap.Could not duplicate.Replaced front end board as a precaution.Re-tested - unit checks ok.Performed electrical safety test.Fcn level: 1416.Expedited qa review: yes.Op = on patient.Unconfirmed.
|
|
Manufacturer Narrative
|
(b)(4).No iabp parts or recorder strips were returned for evaluation at the teleflex (b)(4) facility.The pump was checked by the field service agent and no problem was found.The front end board was replaced for a precautionary.The pump passed the functional check and was returned to service.A device history record (dhr) review is not required.No functional issue was found with the pump.Conclusion: the reported complaint of "trigger loss in autopilot mode" is not able to be confirmed.The pump was check by the field service agent and no issue was found.The pump passed the functional check-out.No parts or recorder strips were returned for evaluation.The cause of the reported complaint could not be determined.
|
|
Event Description
|
It was reported via a field service report: (b)(4).Symptom: loss of trigger in autopilot mode - could not "re-acquire." findings / action taken: ran for 1 hour in all trigger in ecg and ap.Could not duplicate.Replaced front end board as a precaution.Re-tested - unit checks ok.Performed electrical safety test.Fcn level: 1416.Expedited qa review: yes.Op = on patient unconfirmed.
|
|
Search Alerts/Recalls
|