• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLAIR ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLAIR ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FAF07070
Medical Device Problem Codes Positioning Failure (1158); Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/05/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or no images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during deployment of the stent, the endovascular stent graft allegedly failed to deploy from the endovascular graft delivery system.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed with special attention to the subassemblies, material review reports, raw material testing, manufacturing process, and quality control inspection.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.Visual/microscopic inspection: the sample was returned.The safety clip was missing upon receipt.The tuohy borst adapter was disconnected and loose and the two-way stop cock was closed.The outer catheter was noted to be stretched 49m from the strain relief.The stent graft was partially deployed and protruded approximately 9mm from the tip of the outer catheter.Dried fluid was observed between the outer catheter and the stent graft.The deployed portion of the stent graft had the eptfe material peeling from the stuts appearing shredded in nature.It was unknown when this occurred or if the device was flushed with a material that caused eptfe decay.As this was not reported by the user it was considered an incidental finding.Functional/performance evaluation: an attempt to deploy the stent graft was not successful, as the inner catheter could not be moved relative to the outer sheath.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the investigation was confirmed for partial deployment, as the stent graft was returned partially deployed.The root cause could not be determined based upon available information.It was unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) state: indication for use: flair endovascular stent graft is indicated for use in the treatment at the venous anastomosis of eptfe or other synthetic arteriovenous (av) access grafts.Warnings: the stent graft (implant) cannot be repositioned after total or partial deployment.Once partially or fully deployed, the flair endovascular stent graft cannot be retracted or remounted onto the delivery system.Device removal after deployment can only be done with a surgical approach.Precautions: the safety and effectiveness of this device have not been established when used around a tight bend in a looped av graft.It is recommended to access the av graft at the venous side of the av graft or at the apex.Careful attention should be paid to ensure the device is appropriately sized to the actual graft diameter, taking into account any change to the stated graft diameter that may have resulted from previous interventions.The appropriate length device(s) should be selected so that the stent graft extends beyond the stenosis into at least 10 mm of the non-diseased graft towards the arterial inflow and into the non-diseased vein approximately 10 mm beyond the stenosis.During deployment of the stent graft, the entire length of the delivery system should be kept as straight as possible in order to mitigate the potential for distal stent graft misplacement.After full stent graft deployment, wait a few seconds to allow for complete device expansion before removing the delivery system over the guidewire.Stent graft dislodgement has been reported during removal of the delivery system; therefore, careful attention should be paid during this portion of the procedure to prevent such occurrences.In case of placement of two stent grafts (overlap), use the same device diameter in both cases.If a flared device is used to overlap, do not deploy the flared end inside the first stent graft.Ensure a minimum 10 mm overlap of the devices.Potential complications and adverse events: stent graft specific events that could be associated with clinical complications include stent graft misplacement, stent graft migration, stent graft fracture, stent graft kinking, insufficient stent graft expansion and stent graft embolism.Delivery system specific events that could be associated with clinical complications include bond joint failures, detachment of parts, incompatibility with accessory devices, premature deployment, inaccurate deployment, failure to deploy, high deployment forces, delivery system kinking, no visibility under fluoroscopy, inability to track to target location, and blood leakage from delivery system (hemostasis).Stent graft size selection: special care must be taken to ensure that the appropriate size flair endovascular stent graft is selected.The stent graft body diameter should be approximately 1 mm larger than the synthetic av access graft diameter.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during deployment of the stent, the endovascular stent graft allegedly failed to deploy from the endovascular graft delivery system.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLAIR ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6138214
Report Number2020394-2016-01123
Device Sequence Number18990772
Product Code MIH
UDI-Device Identifier04049519008806
UDI-Public(01)04049519008806(17)190617(10)ANAT2359
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P060002/S002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/17/2019
Device Catalogue NumberFAF07070
Device Lot NumberANAT2359
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/07/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/01/2016
Supplement Report FDA Received Date01/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-