Catalog Number 48361234
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/03/2016
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that due to severe sacral tilt, surgeon struggled trialing at l5-s1.Two trial tips were broken and one tip remained inside the trial itself.
 
Additional Manufacturer Narrative
Visual inspection, deice history review, complaint history review, risk assessment.The tip of the returned trial handles were fractured.No relevant manufacturing issues found upon manufacturing history review.The plausible root causes for the reported event are normal wear and tear and user error.
 
Event or Problem Description
It was reported that due to severe sacral tilt, surgeon struggled trialing at l5-s1.Two trial tips were broken and one tip remained inside the trial itself.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIDE TRIAL HANDLE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR  33610
MDR Report Key6138236
Report Number0009617544-2016-00478
Device Sequence Number10918520
Product Code LXH
UDI-Device Identifier07613327269147
UDI-Public(01)07613327269147
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number48361234
Device Lot Number133999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/04/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/01/2016
Supplement Report FDA Received Date01/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-