Medtronic received information that during treatment of an aneurysm located in the cavernous segment of the ica, the pipeline did not open distally as the patient had severe vessel tortuosity.There was also a report of movement during placement, moderate difficult placement/positioning.It was reported that the patient had a proximal loop in ica and a torturous posterior and anterior genuine of ica.During advancement of the pipeline through the marksman at the cavernous segment, the tip of the system at the distal marker appeared to be kinked at a 45 degree angle.This kink was present as the device was advanced up to the m1 segment.During initial delivery the delivery wire at the distal marker appeared to be kinked and the pipeline device was opening in a trumpet shape.There was resistance encountered during advancement of the pipeline within the distal segment of the marksman.After resheathing multiple times, the device exhibited the same trumpet shape and the physician decided to abort the procedure.The device was removed by retracting the marksman and pipeline into the guide catheter and removed from the patient.No patient injury was reported.Post angiography result was normal.The aneurysm was unruptured and saccular.The max diameter was 13 mm and the neck diameter was 10 mm.The distal landing zone was 4.9 mm and the proximal was 4.2 mm.More than 50 percent of the device was deployed, and the device was resheathed less than or equal to 2 times.
|