• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616S
Device Problem Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/22/2016
Event Type  malfunction  
Event Description
It was reported that in a routine inspection, it was noticed that the motor was generating heat.No reported patient injuries.No other complications were noted.
 
Manufacturer Narrative
Complaint of overheating was confirmed.Mdu seized up, failed for motor stall error, and overheated during functional testing.Cause of overheating and errors is a corroded motor/gearbox.The motor/gearbox was removed from the housing.The gearbox was removed from motor and motor tested good.The gearbox was found to be jammed and corroded internally.A review of the device history record was performed which confirmed no inconsistencies.After the evaluation the root cause for the reported issue was determined to be corrosion.There are no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC REPL,MDU, HAND CNTRL, PWRMX
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
5123585706
MDR Report Key6139821
MDR Text Key61537218
Report Number1643264-2016-00259
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 11/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2012
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-