Model Number MICL13.2
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/31/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event problem and evaluation codes: results codes: visual inspection of the returned product found the lens stuck inside an sfc-45 fp cartridge.The cartridge tip was split and a foam tip plunger was stuck inside.There was evidence of dry surgical residue.Conclusions code: based on the complaint history, work order search and the evaluation of the returned product, a specific root cause of the event could not be determined.(b)(4).
 
Event or Problem Description
The reporter stated a 13.2mm micl13.2 implantable collamer lens, -9.0 diopter, became stuck in the loader.There was no patient contact and the backup lens was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key6140268
Report Number2023826-2016-01710
Device Sequence Number1248291
Product Code MTA
UDI-Device Identifier00841542103275
UDI-Public(01)00841542103275(17)180430
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/30/2018
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2016
Initial Report FDA Received Date12/01/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL SFC-45 FP - LOT NUMBER 1330315; FOAM TIP PLUNGER MODEL FTP - LOT NUMBER UNK; INJECTOR MODEL MSI-PF - LOT NUMBER UNK
Patient Age33 YR
-
-