| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/31/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Event problem and evaluation codes: results codes: visual inspection of the returned product found the lens stuck inside an sfc-45 fp cartridge.The cartridge tip was split and a foam tip plunger was stuck inside.There was evidence of dry surgical residue.Conclusions code: based on the complaint history, work order search and the evaluation of the returned product, a specific root cause of the event could not be determined.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter stated a 13.2mm micl13.2 implantable collamer lens, -9.0 diopter, became stuck in the loader.There was no patient contact and the backup lens was used.
|
| |
|
Search Alerts/Recalls
|