DEPUY SYNTHES POWER TOOLS HAND SWITCH F/EPD INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.012 |
Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
(b)(6).
The actual device was returned for evaluation.
Reliability engineering evaluated the device and identified no failure.
Therefore, the reported condition was not confirmed.
An assignable root cause was not determined.
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that the handswitch device had unstable movement; and that it would stop sometimes but would not stop at other times.
According to the reporter, the possibility of radio disturbances was low since an electric knife was not being used.
The event did not occur during surgery.
There was no patient involvement reported.
There were no reports of injuries, medical intervention or prolonged hospitalization.
The date of event was unknown.
All available information has been disclosed.
If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|