| Model Number |
H7493808008250 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/08/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.(b)(4).Returned product consisted of a quantum maverick balloon catheter.The balloon was tightly folded.There was contrast in the inflation lumen.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube shaft was completely separated 61cm from the hub.The fracture faces were oval as if kinked prior to separation.There were numerous hypotube kinks.There was tip damage.Inspection of the remainder of the device presented no other damage or irregularities.Inspection of the remainder of the device presented no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 16-nov-2016.It was reported that crossing difficulties were encountered.The target lesion was located in a coronary artery.A 2.5mm x 8mm quantum¿ maverick¿ balloon catheter was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed hypotube break.
|
| |
|
Search Alerts/Recalls
|