Model Number N/A
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/01/2016
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.
 
Event Description
It was reported that upon opening implant box, it was noticed that the implant had perforated the double plastic wrapping.Another implant was opened.No harm to patient.No debris found.Surgery was not delayed.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: correction: complaint sample was evaluated and the reported event was confirmed.The femoral stem packaging was returned and from the returned package evaluation determined that, the inner and outer cavities were cracked on the cavity flange, outer cavity was damaged and the outer carton box shows the transit damage.Device history record was reviewed and no discrepancies were found.The cracked cavities damage can be caused due to the outside forces during transit or the forces due to the implant.As it's unknown how many times it has been in shipping and it was in the field for a long time, the root cause for the reported issue is attributed to be transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL STEM 12/14 NECK TAPER PLASMA SPRAYED PRESS-FIT CEMENTLESS SIZE 5
Type of Device
HIP, PROSTHESIS
Manufacturer (Section D)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer (Section G)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6143761
MDR Text Key61699765
Report Number0001822565-2016-04520
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PK032726
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date03/31/2018
Device Model NumberN/A
Device Catalogue Number00771100520
Device Lot Number60917333
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-