• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Catalog Number TRI-30M
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/09/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: boston scientific jagwire wire guide, unknown model.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Improper cutting wire orientation can occur if the distal end of the catheter is shaped manually.This sphincterotome catheter is pre-curved and is provided with a pre-curved stylet in the distal tip of the catheter.This obviates the need for manual formation.The instructions for use contain the following comment: ¿note: do not apply manual pressure to tip or cutting wire of sphincterotome to influence orientation, as this may result in damage to device.¿ other factors that can contribute to improper cutting wire orientation include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user to "uncoil and straighten sphincterotome" upon removing device from packaging.The user is then instructed to "carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all tri-tome pc triple lumen sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook tri-tome pc triple lumen sphincterotome.During a bilio-pancreatic endoscopic catheterization, the sphincterotome that was being used was bent and non-functional [tip bent/incorrect cutting wire orientation].The integrity of the device had not been verified before use.The physician took out the sphincterotome, and introduced it again.It was "completely twisted".Another device was used without further problems.
 
Additional Manufacturer Narrative
Concomitant medical products: boston scientific jagwire wire guide, unknown model.Investigation evaluation: our evaluation of the returned device confirmed the report of incorrect cutting wire orientation.During the laboratory analysis, the sphincterotome was advanced through a duodenoscope that is placed in a simulated biliary position.The duodenoscope has an accessory channel that is 4.2 mm in diameter (model number olympus tjf-160v).The catheter exited the endoscope with the cutting wire facing 9 o'clock.The device was then bowed and the cutting wire was facing 6 o'clock (appropriate orientation is approximately 11:00 - 1:00 o'clock).Prior to functional testing, the sphincterotome catheter was subjected to a close visual examination.A discrepancy or anomaly that could have contributed to this observation was not found during our laboratory analysis of the returned product.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Improper cutting wire orientation can occur if the distal end of the catheter is shaped manually.This sphincterotome catheter is pre-curved and is provided with a pre-curved stylet in the distal tip of the catheter.This obviates the need for manual formation.The instructions for use contain the following comment: ¿note: do not apply manual pressure to tip or cutting wire of sphincterotome to influence orientation, as this may result in damage to device.¿ other factors that can contribute to improper cutting wire orientation include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome." the user is then instructed: "carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all tri-tome pc triple lumen sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME
Common Device Name
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key6144482
Report Number1037905-2016-00490
Device Sequence Number710645
Product Code KNS
UDI-Device Identifier00827002225572
UDI-Public(01)00827002225572(17)190922(10)W3774042
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K901443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 02/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberTRI-30M
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date11/09/2016
Device Age2 MO
Event Location Hospital
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/10/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/05/2016
Supplement Report FDA Received Date02/24/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS TJF-Q180V DUODENOSCOPE
Patient Age70 YR
-
-