• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD ARCHITECT TSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD ARCHITECT TSH Back to Search Results
Catalog Number 07K62-35
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/09/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Lot/serial number was manufactured prior to udi compliance date; therefore, only a di is provided an evaluation is in process.A follow up report will be submitted when the evaluation is complete.Note: this is the same patient as reported in in manufacturers report number 3005094123-2016-00054 on concomitant sn (b)(4).
 
Event or Problem Description
The customer reports a falsely depressed tsh result on the architect i2000sr analyzer.The following data was provided from (b)(6) 2016: sid (b)(4) initial 0.016 miu/l, repeat 3.17 miu/l.The result was not reported out of the lab and there was no impact to patient management reported.
 
Additional Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and accuracy testing.Historical performance of the reagent lot was evaluated using world wide data from abbott link.The patient data was analyzed and compared to an established control limit.This evaluation indicated that the patient median result for this lot is within the established control limits and no unusual reagent lot performance was identified.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The product was not available for return.An internal panel was tested with retained kits of the likely cause reagent lot and accuracy testing met all specifications.Based on all available information and abbott diagnostics' complaint investigation, the assay performed as intended and no product deficiency was identified.
 
Additional Manufacturer Narrative
The device evaluation was reassessed and concluded that a malfunction occurred, therefore, the device was not performing as intended, however, no systemic issue or product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TSH
Common Device Name
TSH
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford NA
EI  NA
MDR Report Key6145602
Report Number3005094123-2016-00055
Device Sequence Number10878871
Product Code JLW
UDI-Device Identifier00380740105969
UDI-Public00380740105969
Combination Product (Y/N)N
Initial Reporter CountryNL
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/30/2017
Device Catalogue Number07K62-35
Device Lot Number66208UI00
Was Device Available for Evaluation? Yes
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/14/2016
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date12/05/2016
Supplement Report FDA Received Date12/22/2016
04/21/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT I2000SR ANALYZER; LN (B)(4) SN (B)(4)
Patient Age7 DA
-
-