| Model Number |
ENAV |
| Medical Device Problem Code |
Imprecision (1307)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/09/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient identifier field not sufficient to hold all digits, should read: (b)(6).On 11/22/2016 a medtronic representative noted an attempt to verify navigation system accuracy during an accuracy check, however, did not feel that the unit was accurate.The medtronic representative performed a navigation system check-out, all areas passed.System performed as intended on 11/29/2016 software investigation completed.Findings are that the patient has soft skin and does not appear to be an ideal candidate for tracer.The tracer pattern did not appear posterior and lateral patient anatomy.Software is functioning as designed.- no further issues have been reported.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that, while in a cranial resection procedure, the surgeon alleged an inaccuracy occurred.The inaccuracy was alleged to be 1 centimeter posterior within mach cranial software and was noted once the patient was sterile.The medtronic representative noted the patient had loose skin and the skin was manipulated on 3d model from headphones during mr.Surgical entry point was near ear.No further details regarding this issue were provided.The surgeon completed the procedure with the use of the navigation system.There was no delay of therapy.There was no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
The instructions for use (ifu) that accompanies the device provides guidance for multiple registration methods.
|
| |
|
Additional Manufacturer Narrative
|
|
Device manufacture date updated.
|
| |
|