DEPUY SYNTHES POWER TOOLS ELECTRIC PEN DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.010 |
Device Problems
Corroded (1131); Self-Activation or Keying (1557); Device Inoperable (1663)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.It was further determined that the device had a severely corroded motor and controller due to improper cleaning.The assignable root cause was determined to be due to improper reprocessing, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by (b)(6) that the electric pen drive device was not working when used together with the drill/burr attachment device and the reciprocating saw attachment device.During the pre-repair diagnostics assessment, it was determined that the electric pen drive device ran by itself and the motor/control was severely corroded.It was further determined that the device failed for check direction of rotation, check function with test hand switch and for check function with foot pedal.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|