MAQUET CARDIOPULMONARY AG OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HMOD 70000-USA
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/28/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The product was not available for investigation; therefore manufacturers' laboratory investigation was not possible.No product lot # was provided; therefore it was not possible to check if the specific product is still within capa (b)(4) range (effectiveness check).Furthermore it was not possible to perform a dhr review.Trend search: a trackwise trend search was performed for p/n 70105.3815 failure code d-leakage gas outlet and 7 complaints were found.Three of them are already closed and were referencing to capa (b)(4).A sap trend search was performed for p/n 70105.3815 failure code 0101 leakage gas outlet and 34 additional complaints were found to be referencing to capa (b)(4).Based on this the failure is known and was thoroughly investigated under capa (b)(4).The investigation has identified that the root cause is due to the manufacturing process of the raw material fibers used in the oxygenator.If, during the surface pre-treatment, a system malfunction results in a system results in a system stop, the entire length of the fiber is not properly treated to modify the surface tension.If these untreated fibers are present in the epoxy area of the oxygenator, it is not properly fixed in the epoxy.When this occurs, the fibers are able to ¿shrink out¿ of the epoxy and result in the reported leakage.The raw material manufacturer has initiated steps to repeat the surface pre-treatment to ensure that the proper surface tension is present on the entire length of the fibers.
 
Event Description
According to the customer: "blood leak from exhaust (gas out) port.The product was exchanged during patient treatment.No patient consequences were reported." (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6148614
MDR Text Key61868370
Report Number8010762-2016-00709
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
PMA/PMN Number
K101153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHMOD 70000-USA
Device Catalogue Number70105.3815
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2016
Initial Date FDA Received12/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-