| Catalog Number |
RSS901 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Blood Loss (2597)
|
| Date of Event |
11/04/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Udi number for this product code/lot number combination is not required.The di portion is provided.The actual device was returned to the manufacturing facility for evaluation.The investigation is currently ongoing.A follow up report will be submitted when the investigation is complete, but no later than 30 days from the date that this report was sent.For this reason code 11 was used in the conclusions.The same user facility reported similar events for other devices with the same product code, see mdr's 1118880-2016-00270 and 1118880-2016-00271.All available information has been placed on file in quality assurance at the manufacturing facility for appropriate tracking, trending and follow-up.
|
| |
|
Event or Problem Description
|
|
The user facility reported a leak on the involved device.Follow up communication with the user facility confirmed the following information: a leaky valve was reported during a case; and there was no patient injury or medical intervention required.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned sample was visually inspected.The valve was dislodged and off center damage was observed on the valve.An evaluation of the retention sample from the reported and surrounding lots was conducted.There were no visual anomalies noted during visual inspection and all samples met the leak test specification.The valve was present and functioned as intended.A review of the device history record of the product code/lot# combination was conducted with no relevant findings.There is no evidence that this event was related to a device defect or malfunction.Although the exact cause cannot be definitively determined, based on the available information it is likely that the dilator was inserted off center which caused the valve to dislodge resulting in the leakage reported.The potential for such an event is addressed in the instruction-for-use (ifu) with statements such as the following: "insert the dilator into the valve center of sheath.Forced dilator insertion missing the valve center may cause valve damage, resulting in blood leakage." all available information has been forwarded to the manufacturing facility for appropriate tracking, trending and follow up.
|
| |
|
Event or Problem Description
|
|
Additional information provided on (b)(6) 2016 confirmed the following; blood loss was less than 250 ml; and the sheath was replaced with another sheath.
|
| |
|