• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - CORK CRE¿ FIXED WIRE; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - CORK CRE¿ FIXED WIRE; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558380
Device Problems Loss of or Failure to Bond (1068); Deflation Problem (1149)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.(b)(4).
 
Event Description
It was reported to boston scientific corporation that a cre fixed wire balloon was used during an esophagogastroduodenoscopy (egd) with balloon dilation procedure on an unknown date.According to the complainant, during preparation, a piece of plastic that was believed to be the balloon bond was noted to look broken.During the procedure, the balloon was able to still be dilated.At the conclusion of the procedure, it is believed the balloon bond prevented the deflation of the balloon, so the device was removed from the endoscope by cutting off the top of the balloon.The procedure was completed at this time.There were no patient complications reported as a result of the event.The patient¿s condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE¿ FIXED WIRE
Type of Device
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI 
Manufacturer (Section G)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI  
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6149435
MDR Text Key61616541
Report Number3005099803-2016-03803
Device Sequence Number1
Product Code KNQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Model NumberM00558380
Device Catalogue Number5838
Device Lot Number0019627984
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/24/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-