| Model Number |
H7493808015400 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/11/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 90% stenosed target lesion was located in the left anterior descending (lad) artery.A 4.0mm x 15mm quantum¿ maverick¿ balloon catheter was selected for dilatation.However, during the procedure, the balloon shaft fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The balloon was tightly folded.There was contrast in the inflation lumen.There was blood in between balloon folds.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube shaft was completely separated 57.5 cm from the hub.The fracture faces were oval as if kinked prior to separation.There were numerous hypotube kinks.There was no evidence of any material or manufacturing deficiencies contributing to the damage.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was further reported that the device was completely removed from the patient.
|
| |
|