| Catalog Number |
9735017 |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/10/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Return requested for suspect malleable angle tip suction (b)(6) 2016.No parts have been received by manufacturer for analysis.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that, while in a ear, nose & throat (ent) procedure, malleable suction verification issue.The instrument was functioning normally, however,unexpectedly stopped tracking.Another malleable suction was opened and was used to continue the procedure without issue.The surgeon completed the procedure with the use of the navigation system.Delay in therapy was less than one hour.There was no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Usage or device and labeled for single use? values provided.The hardware investigation of the returned suction found that the reported event was related to an electrical issue.As reported, the suction was not identified and would not track when connected to a known good system.Coil #3 was nonfunctional and caused the issue.The reported event was confirmed.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
| |
|