| Model Number |
N/A |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The event is currently under investigation.
|
| |
|
Event or Problem Description
|
|
It was reported the performer introducer valve was leaking during the procedure.The device was exchanged during the procedure.The patient had no adverse consequence due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation - evaluation: a review of the complaint history, drawings, device history record, documentation, manufacturing instructions and quality control of the returned device was conducted during the investigation.The complaint device was not returned therefore, no physical examinations could be performed; however, a document based investigation evaluation was performed.No images were available and a review of the product was not performed as the device was not returned for evaluation.The device history record was reviewed and no non-conformances were noted.A search revealed this complaint to be the only reported complaint associated to the complaint lot number 6968919.Although the devices were not returned for evaluation, the root cause was determined to be manufacturing related.Measures have been previously initiated to address this issue.A quality engineering risk assessment has determined that no further action is required as it concluded that its occurrence is unlikely to lead to a serious health consequence to the patient.We will continue to monitor for similar complaints.
|
| |
|
Event or Problem Description
|
|
It was reported the performer introducer valve was leaking during the procedure.The device was exchanged during the procedure.The patient had no adverse consequence due to this event.
|
| |
|
Search Alerts/Recalls
|