• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZKB00
Medical Device Problem Codes Material Puncture/Hole (1504); Device Damaged by Another Device (2915)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Information (3190)
Date of Event 11/08/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was punctured by the inserter during insertion into the eye.Information for the inserter or cartridge was not provided.No further information was provided.
 
Additional Manufacturer Narrative
Additional info: additional information was received and it was learned that the intraocular lens (iol) was fully inserted into the patient's eye and it was removed and replaced, during the same surgical procedure.No medical or surgical intervention was required.(b)(4).The inserter and cartridge model and lot number were also provided.Concomitant products: cartridge 1mtec30, lot cb36700; injector dk7796, lot 146.All pertinent information available to abbott medical optics has been submitted.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6155675
Report Number9614546-2016-00663
Device Sequence Number1053072
Product Code MFK
UDI-Device Identifier05050474560147
UDI-Public(01)05050474560147(17)200120
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date01/20/2020
Device Model NumberZKB00
Device Catalogue NumberZKB00U0230
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/09/2016
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date12/08/2016
Supplement Report FDA Received Date01/31/2017
04/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INSERTER & CARTRIDGE: UNKNOWN MODELS, LOT/SN#
-
-