• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZKB00
Medical Device Problem Codes Material Puncture/Hole (1504); Device Damaged by Another Device (2915)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Information (3190)
Date of Event 11/08/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was punctured by the inserter during insertion into the eye.Information for inserter or cartridge was not provided.No further information was provided.
 
Additional Manufacturer Narrative
Additional info: additional information was received and it was learnt that the intraocular lens (iol) was fully inserted into the patient's eye and it was removed and replaced, during the same surgical procedure.The inserter and cartridge model and lot number were also provided.Concomitant medical products has been updated: cartridge 1mtec30, lot cb36700; injector dk7796, lot 146.All pertinent information available to abbott medical optics has been submitted.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6155689
Report Number9614546-2016-00664
Device Sequence Number19089389
Product Code MFK
UDI-Device Identifier05050474560178
UDI-Public(01)05050474560178(17)200311
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date03/11/2020
Device Model NumberZKB00
Device Catalogue NumberZKB00U0245
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/09/2016
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date12/08/2016
Supplement Report FDA Received Date01/27/2017
04/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INSERTER & CARTRIDGE: UNKNOWN MODELS, LOT/SN#
-
-