|
(b)(4): ball tip 1mm probe, lot # 0211194199, manufactured date ¿ apr/28/2016, expiration date ¿ apr/26/2024, 510(k) # k934426, (b)(4).The product was returned for analysis.Device evaluation anticipated but not yet begun.
|
|
Analysis of emg tube (product number 8229707, lot number 0210611979) one un-sealed sample, part number 8229707, from lot number 0210611979 was received.Visually, there was no damage or anomalies in the construction of the device that would have resulted in the reported event: ¿the tube and probe were not detecting nerves / not responding.¿ the resistance of the electrodes from end to end shall be less than 200 ohms and the actual measurement of each circuit are as follows; red 22 ohms, red [w/white band] 19 ohms, blue 25 ohms, and blue [w/white band] 18 ohms.There were no open circuits and no out of specification condition of the main tube electrodes.There were no shorts between circuits and no intermittent behavior while manipulating the tube and wires.When viewed under magnification, there were no cracks in the electrode contacts.The electrode contacts / pads were aligned with the curvature of the tube and midline.Per the xvi a continuity test is performed prior to final packaging.The ifu provides contraindications and warnings that identify reasons why a user may experience reduced or eliminated emg responses to direct or passive neural stimulation such as; paralyzing agents / lubricants / sprays, neuromuscular fatigue from prolonged or repeated exposure to electrical stimuli, deep anesthesia, which may suppress neuroelectrical activity, insufficient contact between the electrodes and vocal cords due to misplacement or using a tube size that is too small, and displacement during the procedure when rotating, flexing, or extending the patient¿s head and neck.There was no fault found and no evidence of improper manufacturing, therefore manufacturing has been ruled out as a likely cause.There was no evidence of improper manufacturing and therefore has been ruled out as a likely cause.The device condition was in specification / performed as expected, as it relates to the complaint.Analysis of ball tip probe (product number 8225275, lot number 0210611979) one un-sealed sample, part number 8225275, from lot number 021119419 was received.There was no damage to the packaging to indicate a cause.Visually, there was no damage or anomalies in the construction of the device that would have resulted in the reported event.The probe was installed into a nim system and functionally tested at 1.0ma stimulating current with a 100uv threshold; there was a 1.02 stimulation return and a waveform / tone around 200uv indicating current delivery; the probe performed as intended.There was no fault found or evidence of improper manufacturing therefore has been ruled out as a likely cause.There was no evidence of improper manufacturing and therefore has been ruled out as a likely cause.The ifu provides multiple reasons that will significantly reduce, if not completely eliminate, emg responses to direct or passive nerve stimulation as a result of use error.The device condition was in specification / performed as intended, as it relates to the complaint.(b)(4).
|