Merge technical support remotely accessed the site's server and found an error message indicating that the pdm (patient data module) was disconnected from the hemo monitor.Technical support asked the customer if there were any missing or broken screws on the pdm.The customer stated that there was tape on the cable that leads into the pdm.Subsequently, replacement hardware was shipped to the customer.The customer confirmed that the problem was resolved after the hardware was replaced.However, six (6) days later the customer called back and reported that the hemo monitor froze three (3) times during a procedure.The hemo monitor was also rebooted.Reference mdr #21839216-2016-00812 (b)(4).Device labeling (v10 user manual) addresses such an occurrence with statements such as, "inspect overall physical condition of the system components, peripherals, and interconnecting cables.Perform any corrective actions required.".
|
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that problems were experienced with the hemo monitor where intermittent freezing occurred during two (2) procedures.This required the user to reboot the entire hemo system resulting in a loss of patient monitoring.The reported delay was ~5 minutes.With merge hemo not capturing physiological data, there is a potential for delay of treatment that could cause harm to the patient.The customer reported that the procedure was completed successfully once the hemo monitor was rebooted.(b)(4).
|