SYNTHES OBERDORF COMPACT AIR DRIVE II (CAD II); INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 511.701 |
Device Problems
Mechanical Problem (1384); Power Problem (3010)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/15/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to improper handling, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the compact air drive device had lack of power.During the pre-repair diagnostics assessment, it was determined that the motor power was too low.It was further determined that the device had lack of power, ¿body use¿ and lack of lubrication.It was further determined that the device failed for general condition, marking and labeling, check for untrue running, cheek for excessive noise, check the power with test bench: min.11 0 to 160 w and for check starting behavior.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|