ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Failure to Convert Rhythm (1540)
Patient Problem Death (1802)
Event Date 11/01/2016
Event Type  Death  
Manufacturer Narrative
Device evaluation of monitor sn (b)(4) and electrode belt sn (b)(4) has been completed.During the incoming functional testing, a 1hz simulated normal sinus rhythm signal was applied to the ecg electrodes, followed by a 5hz simulated treatable arrhythmia signal which verified proper performance of the detection algorithm.During the transition to the 5hz signal, the device was confirmed to properly enter into a treatment sequence which includes a verification of the tactile vibration alarm, audio messaging, and siren alarms, as well as a test of the pulse delivery circuitry.The pulse delivery circuitry test verified proper charging of the high voltage capacitors and proper delivery of five full energy 150j biphasic pulses.The functional testing confirmed proper response button functionality, ecg acquisition, detection algorithm performance, and pulse delivery functionality.There is no indication of a product malfunction.
 
Event Description
A us distributor contacted zoll to report that the patient passed away on (b)(6) 2016.It was reported the patient was in the icu at a medical facility.The patient received eleven appropriate treatments between 09:41:18 and 09:50:01 during vt/vf.Following each of the treatments the rhythm failed to convert.Following each treatment the patient's rhythm remained in vt/vf.The device was shut down on (b)(6) 2016 at 09:50:25 am.There is no device malfunction that caused or contributed to the patient death.The device correctly detected vt/vf and delivered eleven treatments.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
michelle hilke
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key6159215
MDR Text Key62073979
Report Number3008642652-2016-08969
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial
Report Date 12/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2016
Date Manufacturer Received11/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age65 YR
-
-