• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TYRX, INC. AIGIS-R LARGE; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TYRX, INC. AIGIS-R LARGE; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number NMRM6133
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/07/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation conclusion: lot# 16h15753 - the particle was located on the interior surface of the envelope.The particle had a dark and glassy appearance in the optical images.The particle was about 1mm in length.The sample was sent to the lab to determine the origin of the foreign matter.The unknown residue was thoroughly characterized by infrared spectroscopy and tof-sims.The results show the unknown material from the field return is similar to material from the spraying, and is composed of a mixture of thk afa grease and api coating (tyrosine polymer, minocycline and rifampin).A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the implant procedure the absorbable envelope was opened and found to have a visible anomaly on the envelope, a black spec.The envelope was inspected before going to the sterile field by the implanting physician who determined it should be discarded.Another envelope was implanted.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
The device envelope was returned and analyzed.Analysis of the returned product indicated an issue with the device envelope appearance.There was a particle located on the interior surface of the envelope.The particle had a dark and glassy appearance in the optical images.The particle was about 1mm in length.The sample was sent to the lab to determine the origin of the foreign matter.The unknown residue was thoroughly characterized by infrared spectroscopy and time-of-flight secondary ion mass spectrometry (tof-sims).The results suggest the unknown material from the field return is similar to material from the spraying/drying oven, and is composed of a mixture of thk afa grease and active pharmaceutical ingredient (api) coating.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIGIS-R LARGE
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
TYRX, INC.
1 deer park dr.
suites g, e
monmouth jct NJ 08852
MDR Report Key6164742
Report Number3005619263-2016-00055
Device Sequence Number18919903
Product Code FTL
UDI-Device Identifier00643169624795
UDI-Public00643169624795
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K150291
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2016
Device Model NumberNMRM6133
Device Catalogue NumberNMRM6133
Device Lot Number16H15753
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2016
Initial Date Received by Manufacturer 09/07/2016
Supplement Date Received by ManufacturerNot provided
10/10/2016
Initial Report FDA Received Date12/09/2016
Supplement Report FDA Received Date04/06/2017
09/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 YR
-
-