• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Failure to Prime (1492)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a prime (unable to prime) issue.There was no indication that the product caused or contributed to an adverse event.This issue is being reported because it may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Manufacturer Narrative
Follow-up #1: date of submission 02/10/2017.Device evaluation: the device has been returned and evaluated by product analysis on 01/20/2017 with the following findings: a review of the black box showed multiple loss of prime warnings with low non zero and zero force, and no cartridge warnings.The rewind, load, and prime steps were performed successfully.The pump was exercised for 24 hours and the loss of prime and load step malfunctions were not duplicated.The force calibration testing passed.The pump opened to find no defect found.Unrelated to the original complaint, the battery compartment was cracked from the case seal traveling to the grip pad, and at the case seal to the left side of the grip pad.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6164962
MDR Text Key62333918
Report Number2531779-2016-33323
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10840406100051
UDI-Public0110840406100051
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 11/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received11/16/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age54 YR
Patient Weight210
-
-