|
The event was reported by a customer from usa: "per clinical department, the issue is slit on tubing at slider clamp area- 12000-000-0027- ap431-01 - infusion set with vented/non-vented drip chamber,back check valve, 2 needleless y-site injuries or adverse event: no; item: (b)(4); quantity affected: 1 each; serial/lot number: n/a; po #: (b)(4); are any samples available for return? no.Thirty (30) cc leaks from tubing to floor in pt room, no human harmed, splash of chemo coloring unto rn shoe.Death/serious injury: no, human harm: no, delay in therapy: no, medical intervention needed: no.".
|