Model Number 631608P
Medical Device Problem Code Hole In Material (1293)
Health Effect - Clinical Code No Information (3190)
Date of Event 11/21/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Method: actual device not evaluated - the device was not returned to vascutek as it remains in situ within the patient.Manufacturing review- a complete review of the process, manufacturing and qc records was carried out which confirmed there were no issues with the manufacture of the device.Process review - a complete review of the process, manufacturing and qc records was carried out which confirmed there were no issues with the manufacture of the device.Sterilisation review- a review of the sterilisation records for this device was carried out which confirmed that the device was sterilised to specification.Results: no failure detected -no issue with the device could be identified from the investigation carried out by vascutek.A complete review of the process, manufacturing and qc records was carried out which confirmed there were no issues with the manufacture of the device.No results available since no evaluation performed - the device was not returned to vascutek as it remains in situ within the patient conclusion: unable to confirm complaint - vascutek are unable to confirm the complaint form the investigation carried out.Device not returned - the device was not returned to vascutek as it remains in situ within the patient.A review of previous complaints vs sales was carried out and the complaint rate for leakage of this product type is (b)(4).No adverse trends were identified form the review carried out.No further action will be taken at this time; however, the issue will be tracked and trended as part of the routine complaints monitoring process.If an adverse trend develops action may be taken at that time.Vascutek now considers this event closed.
 
Event or Problem Description
The event was described to vascutek as; when the clamp was released, blood leaked from the graft leg on the distal side 1 to 2 cm from the branch.Two bleeding points of needle hole size were found and sutures were applied.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GELSOFT PLUS BIFURCATE
Common Device Name
GELSOFT PLUS BIFURCATE
Manufacturer (Section D)
VASCUTEK LTD.
newmains avenue
inchinnan business park
renfrewshire, scotland PA4 9 RR
UK  PA4 9RR
MDR Report Key6165663
Report Number9612515-2016-00032
Device Sequence Number509968
Product Code DSY
UDI-Device Identifier05037881115528
UDI-Public05037881115528
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K955230
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/12/2016
Report Date (Section F) 11/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2021
Device Model Number631608P
Device Catalogue Number631608P
Device Lot Number385316-8983
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Device Age5 MO
Event Location Hospital
Date Report to Manufacturer11/28/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2016
Initial Report FDA Received Date12/12/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-