MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEL06100
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/14/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide further patient details.
 
Event or Problem Description
It was reported that during a treatment of an aneurysm in the sfa the endovascular stent graft failed to deploy.The access site was gained contralateral.The endovascular stent graft system was removed and another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.On the basis of the evaluation of the returned device, the reported deployment failure could be confirmed.The stent graft was found to be loaded in the system.The outer sheath was found to be elongated indicating that increased friction must have affected on the delivery system during the attempt to deploy the stent graft.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This kind of event may be related to a difficult vessel anatomy or a challenging stent placement site.Reportedly, the tracking path was tortuous; the target anatomy was neither tortuous nor calcified.Insufficient flushing of the device may be another contributing factor to this type of event.On the basis of the information available and the evaluation of the sample returned, a definite root cause for the reported event could not be determined.The ifu indicates that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." the reported application represents an off-label use of the device.The device is indicated for use in the venous outflow of hemodialysis patients dialyzing by either an arteriovenous (av) fistula or av graft.The safety and effectiveness of the device when placed across an aneurysm or a pseudoaneurysm has not been evaluated.
 
Event or Problem Description
It was reported that during a treatment of an aneurysm in the sfa the endovascular stent graft failed to deploy.The access site was gained contralateral.The endovascular stent graft system was removed and another device was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
MDR Report Key6166121
Report Number9681442-2016-00325
Device Sequence Number675523
Product Code PFV
UDI-Device Identifier04049519008066
UDI-Public(01)04049519008066(17)180731(10)ANZG3614
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2018
Device Catalogue NumberFEL06100
Device Lot NumberANZG3614
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/14/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/12/2016
Supplement Report FDA Received Date01/25/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
Patient Weight65
-
-