|
Customer complaint data was reviewed and no adverse trends were identified.Accuracy testing was completed to evaluate the assay performance using a file kit of reagent lot 02416a000.All replicates met acceptance criteria and the testing passed.A study was reviewed "performance characteristics of six intact parathyroid hormone assays".Sonia l.La'ulu, and william l.Roberts, md, phd.Am j clin pathol 2010;134:930-938, 2010.This study looked at the performance characteristics of 6 intact parathyroid hormone assays, including the architect which was the comparison method.For method comparison, serum and edta plasma samples were tested by all methods.Overall the study found good correlation for all methods, but did indicate there was significant bias between methods.The study also concluded that there are many factors that potentially influence variability for pth measurements, including the method used, pth source (synthetic vs endogenous), population evaluated, vitamin d status, preanalytic variables such as sample matrix, and storage time and temperature.The study assessed some of these factors that can produce variability and confirmed that there is variability between pth assays.In conclusion the study indicated that standardization efforts are warranted and assay-specific decision limits are required.In addition, technical bulletin tb 1000-2016 was reviewed which indicates that the architect intact pth calibrators and controls will have a reduced expiration date (shelf life) to address calibrator instability.A study performed in april 2014 using abbott intact pth calibrators with reduced dating supports the return of product performance to product claims.Similar overall performance to a comparison assay was observed in 2014 as was observed in 2007.The % bias was determined to be 30.2% (95% confidence interval, -6.4 - 66.8) in stat method and 26.2% (95% confidence interval, -9.4 -61.7) in routine method.The architect intact pth assay package insert was reviewed and was found to adequately address the issue.Based on this investigation a product deficiency was not identified.In addition, there is not enough information to reasonably suggest a malfunction had occurred since limited troubleshooting was performed.Additionally, the product claim indicates a positive bias to the comparison assay.No additional product issues were identified; no further investigation is required.
|