• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM; BIORESORBABLE DRUG ELUTING SCAFFOLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM; BIORESORBABLE DRUG ELUTING SCAFFOLD Back to Search Results
Catalog Number 1235250-18
Health Effect - Clinical Codes Angina (1710); Non specific EKG/ECG Changes (1817); Cardiac Enzyme Elevation (1838); Thrombosis (2100)
Date of Event 11/24/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of angina and thrombosis, as listed in the bioresorbable vascular scaffold (bvs) system, absorb, instructions for use, are known adverse events associated with the use of a coronary scaffold in native coronary arteries.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the patient presented with segment elevation myocardial infarction (stemi) on (b)(6) 2016.The procedure was to treat a lesion in the mid left circumflex artery with mild tortuosity and chronic total occlusion.A 2.5x15mm non-abbott balloon catheter was inflated to nominal pressure to predilate the lesion.A 2.5x18mm absorb gt1 scaffold delivery system (sds) was implanted without issue.A 2.75x15mm non-abbott balloon catheter was inflated to nominal pressure to post dilate the scaffold.Twenty six hours later and while still in the hospital, the patient experienced chest pressure, st elevations and an increase in troponin.Optical coherence tomography (oct) was used to confirm that the area around the scaffold was cloudy due to thrombosis.A non-compliant balloon catheter was used to successfully treat the thrombosis in the scaffold.The procedure was ended.Following the case the patient was administered a heparin drip for 24 hours in addition to aspirin and effient.The patient was discharged four days later on november 29, 2016.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Common Device Name
BIORESORBABLE DRUG ELUTING SCAFFOLD
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6167892
Report Number2024168-2016-08808
Device Sequence Number1418681
Product Code PNY
UDI-Device Identifier08717648212963
UDI-Public(01)08717648212963(17)170821(10)6082261
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P150023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/21/2017
Device Catalogue Number1235250-18
Device Lot Number6082261
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2016
Initial Report FDA Received Date12/12/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age61 YR
-
-