(b)(4).The field service technician evaluated the device and found that the cupola fork was damaged and found dust metalon the spring arm.Maquet determined that the device failed to meet its specification due to a collision with another device.Additionally,the device was directly involved with the reported incident however was not being used to treatment or diagnosis of the patient when the event occurred.Per the labeling, users should "check the lightheads for chipped paint, impact marks and any other damage", on daily basis.
|
Customer reported that, during cleaning, the cupola fork was damaged between the fork and the outer ring of the spring arm, and some dust metal fell off.There were no injuries reported.(b)(4).
|