| Model Number |
9520 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/12/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Quest medical, inc.Has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Quest medical, inc.Defers to the patient's physician regarding medical history.
|
| |
|
Event or Problem Description
|
|
The product distributor reported an issue encountered by the clinician while using the intravascular administration extension set.The report details stated that "air was entering thru distal end of manifold." follow-up with the complainant found that there were no patient complications and another device was used instead.The device was returned to the manufacturer for evaluation.``.
|
| |
|
Additional Manufacturer Narrative
|
|
The alleged complaint condition could not be duplicated in the sample device nor in devices from the same lot tested.It is possible that the clinician did not prime the side ports first then the main central port.
|
| |
|