• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEM14040
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/09/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide further patient details.
 
Event or Problem Description
It was reported that the stent graft failed to deploy during treatment of a central vein stenosis in the left brachiocephalic vein.Reportedly, a lot of resistance was encountered during the attempt to deploy the system.The delivery system was removed without issues.Another device from the same brand was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.As part of all complaint investigations, an attempt was made to evaluate the subject device as well as the device usage.In this case, no sample or image was provided for evaluation.Therefore, the reported event could not be reproduced.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This kind of event may be related to a difficult vessel anatomy or a challenging stent placement site.Reportedly, the lesion had been pre-dilated but there were difficulties in advancing the system through the incision during the first attempt without using an introducer sheath.Then an introducer sheath was used and the system could be successfully advanced to the lesion site.Insufficient flushing of the device may be another contributing factor to this type of event.On the basis of the information available and as no sample was returned for evaluation, a definite root cause for the reported event could not be determined.The ifu states that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure" and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." the ifu indicates that an introducer sheath with appropriate inner diameter is required for the procedure.Furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
 
Event or Problem Description
It was reported that during treatment of a central vein stenosis in the left brachiocephalic vein, the stent graft failed to deploy.Reportedly, a lot of resistance was encountered during attempt of the pin and pull the system.The delivery system was removed without issues.Another device from the same brand was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
MDR Report Key6170043
Report Number9681442-2016-00328
Device Sequence Number13720868
Product Code PFV
UDI-Device Identifier04049519008684
UDI-Public(01)04049519008684(17)190401(10)ANAR0111
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2019
Device Catalogue NumberFEM14040
Device Lot NumberANAR0111
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/16/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/13/2016
Supplement Report FDA Received Date12/22/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age38 YR
Patient Weight82
-
-