| Catalog Number |
9734734 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Inoperable (1663); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/15/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was not made available from the site.Device lot number not available.Device manufacturing date is dependent on lot number, therefore, unavailable.Return requested.Replacement navlock universal green tracker shipped to site 11/16/2016.Return requested.Replacement tap 3.75 shipped to site 11/18/2016.No parts have been received by manufacturer for analysis.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that, while in a spine procedure, the site's green navlock tracker and their 3.75 solera tap became fused together when the surgeon was using these devices with a powerease drill.The surgeon stated that his glove may have gotten caught between the tap and the tracker, as well.The surgeon used a different tracker and a different tap combination to continue the procedure.No further details regarding this issue, or specifically when it occurred, were provided.The surgeon completed the procedure with the use of the navigation system.There was no delay of therapy.There was no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Device lot number now provided.Device manufacturing date now provided.Investigation of returned suspect navlock tracker and tap finds that the reported problem was confirmed.Tap seized up in the navlock tracker and was unable to remove the tap from the navlock tracker to do more of a further investigation.The reported issue was confirmed to be caused by a mechanical issue.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
| |
|
Search Alerts/Recalls
|