U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 HI-FLOW KIT; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 HI-FLOW KIT; NRY Back to Search Results
Catalog Number 5MAXACE068KIT
Medical Device Problem Codes Material Separation (1562); Stretched (1601); Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/15/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Results: the penumbra system ace 68 reperfusion catheter (ace 68) was kinked approximately 3.5 cm from the hub, fractured approximately 56.0 cm from the hub and ovalized approximately 120.0 cm from the hub.Conclusions: evaluation of the returned device revealed it was fractured.This damage likely occurred due to forceful retraction of the ace 68 through an rhv that was not sufficiently opened.Further evaluation revealed the ace 68 was kinked in the proximal shaft and ovalized in the distal shaft.These damages may have occurred due to forceful handling or manipulation of the ace 68 during use.The non-penumbra stent retriever and rhv mentioned in the complaint were not returned for evaluation.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the internal carotid artery (ica) using a penumbra system ace 68 hi-flow kit (kit).During the procedure, while following protocol, the physician removed the other manufacturer's stent retriever and then attempted to remove the penumbra system ace 68 reperfusion catheter (ace 68).While retracting the ace 68, the physician did not mention experiencing any resistance; however, the ace 68 became stuck in the rotating hemostasis valve (rhv) and began to stretch then separate.The physician believed that the rhv was too tight.Therefore, the rhv was loosened and the ace 68 was able to be removed intact.The procedure was then completed using a new ace 68.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ACE 68 HI-FLOW KIT
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6170499
Report Number3005168196-2016-01803
Device Sequence Number13807614
Product Code NRY
UDI-Device Identifier00814548016597
UDI-Public00814548016597
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K161640
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date09/15/2019
Device Catalogue Number5MAXACE068KIT
Device Lot NumberF72504
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2016
Initial Report FDA Received Date12/13/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
-
-