ENCORE MEDICAL, L.P. RSP SHOULDER; RSP, PRIMARY HUMERAL MONOBLOCK, 8MM X 108 Back to Search Results
Catalog Number 510-00-008
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 11/14/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to a superficial infection; the humeral stem, liner and glenosphere were removed.The surgeon washed out the wound and reinstalled new components.
 
Additional Manufacturer Narrative
The reason for this revision surgery was a superficial infection.The previous surgery and the revision detailed in this investigation occurred 61 days apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) and product complaint report (pcr) database records show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Common Device Name
RSP, PRIMARY HUMERAL MONOBLOCK, 8MM X 108
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key6171429
Report Number1644408-2016-00954
Device Sequence Number730191
Product Code KWS
UDI-Device Identifier00888912024815
UDI-Public(01)00888912024815
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K100741
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/15/2022
Device Catalogue Number510-00-008
Device Lot Number979F1255
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/14/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/13/2016
Supplement Report FDA Received Date01/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-32-101,LOT 862C2195; 509-00-032,LOT 377P1151
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR