PENUMBRA, INC. INDIGO SYSTEM SEPARATOR 3; DXE Back to Search Results
Catalog Number SEP3
Medical Device Problem Codes Kinked (1339); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/17/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2016-01822.The hospital disposed of the device.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure using an indigo system aspiration catheter 3 (cat3) and an indigo system separator 3 (sep3).During the procedure, the physician experienced a little resistance while advancing the sep3 through the cat3 during use.Upon removal to check on the progress of the procedure, the physician noticed the sep3 and the cat3 had become kinked.Therefore, the procedure was completed using a new sep3, a new cat3, and the same non-penumbra sheath.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM SEPARATOR 3
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6171795
Report Number3005168196-2016-01821
Device Sequence Number1110493
Product Code DXE
UDI-Device Identifier00814548013312
UDI-Public00814548013312
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/02/2018
Device Catalogue NumberSEP3
Device Lot NumberF63407
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2016
Initial Report FDA Received Date12/13/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-