Programming data was reviewed.The impedance was within normal limits throughout the available history.The charge consumed was too low for the battery voltage to be at 25%.Longevity tables estimated between 9.5 and 6.8 years at 33% and 50% duty cycles, respectively, from beginning of life until ifi=yes.Since the battery was already at 25% after less than 1 year, the device appears to be depleting more quickly than expected.The premature battery depletion was most likely due to the manufacturing method used.
|
Generator product analysis was completed.The reported premature end of life was duplicated in the product analysis, or pa, lab.The vns generator, as received, would not interrogate and, therefore, the data could not be downloaded from the generator and the diagnostic vbat calculation, diagnostics, the final configuration r-wave test, and fet could not be performed.With the generator case removed and the battery still attached to the printed circuit board assembly (pcba), the battery measured 0.176 v, confirming an end of service, or eos, condition.The pcba was subjected to postburn electrical tests and failed several electrical tests.Sandpaper was used to remove the observed contaminates from the trimmed edge of the pcba.The observed contamination on the trimmed edge suggested probably electrical paths established between the copper edges of the trimmed edge of the pcba, contributing to the supply current conditions.Remaining residual material on the pcba edge after test tab removal during manufacturing resulted in increased out of specification current consumption for both standby and pulsing modes.This may have been a contributing factor for the reported premature end of life.
|