A review of the manufacturing lot history and sterilization records was conducted.All in-process specifications and release criteria were met, including testing for suture retention (course and wale) and ball burst conducted on the mesh at incoming.Fourier-transform infrared spectroscopy testing, in-process visual inspection of the cured coated mesh/panels, coating density of the cured coated panels, and seal strength testing on both the pre-and-post-sterile atrium-applied packaging seals were also performed, with all requirements being met.Clinical evaluation: c-qur¿ mosaic mesh has a coating derived from biological oil composed of fatty acids, lipids and glycerides (bao bio-absorbable oil coating).It has a coating that is absorbable by the body and if in the event it does flake or come off during a procedure it would cause no harm to the patient, the mesh would then become the equivalent of bare mesh.A physician may determine a product is not suitable for an intended procedure at any time and for any reason he deems significant enough to compromise patient outcomes.In the event the physician has a change of plan during a procedure it would result in a delay in treatment.The instructions for use state under adverse reaction, that complications may occur with the use of any surgical mesh include, but are not limited to, mechanical disruption of the tissue and/or mesh material, possible adhesions when placed in direct contact with the viscera (intestines) and organs.
|