Model Number MERGE HEMODYNAMICS 10.0.1
Device Problems Application Program Problem (2880); Blocked Connection (2888)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2016
Event Type  malfunction  
Manufacturer Narrative
Merge technical support, in communication with the customer, determined that coiled cables had not been replaced in two (2) of the site's cath labs.The pdm (patient data module) was found to be faulty as well.Subsequently, replacement hardware was shipped to the customer on 14nov2016.The customer confirmed that the problem was resolved after the hardware was replaced.Device labeling (hemo v10 user manual) addresses such an occurrence with statements such as, "inspect overall physical condition of the system components, peripherals, and interconnecting cables.Perform any corrective actions required.".
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that problems were experienced with the hemo monitor where the device froze three (3) times during a procedure.This required the user to reboot the entire hemo system resulting in a loss of patient monitoring.With merge hemo not capturing physiological data, there is a potential for delay of treatment that could cause harm to the patient.The customer reported that the procedure was completed successfully once the hemo monitor was rebooted.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key6176765
MDR Text Key62544226
Report Number2183926-2016-00812
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 10.0.1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/14/2016
Initial Date FDA Received12/15/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-