Model Number 25-092-42-91
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/21/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Doctor was inserting the screw into the bone when the head of the screw broke off.A portion of the screw was left in the bone.
 
Additional Manufacturer Narrative
An investigation was performed using visual and stereo microscope inspection of the screw head returned.Tensile cracks were observed under the stereo microscope.Further observation determined there were no indications of material or manufacturing defects.The device history records could not be reviewed since no lot number was provided.The results of the investigation conclude that the root cause for the breakages were due to mechanical overload on the device.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM  78570
MDR Report Key6177571
Report Number9610905-2016-00051
Device Sequence Number862716
Product Code HRS
UDI-Device Identifier00888118037268
UDI-Public(01)00888118037268
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K083432
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number25-092-42-91
Was Device Available for Evaluation? Yes
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/22/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/15/2016
Supplement Report FDA Received Date03/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age35 YR
-
-