• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493808008400
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/29/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The 90% stenosed, 18mm x 3.5mm target lesion was located in the severely tortuous and moderately calcified right coronary artery.A 4.0mm x 8mm quantum¿ maverick¿ balloon catheter was advanced to dilate the lesion.However, it was noted that the shaft had fractured.The device was completely removed from the patient's body and the procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Additional Manufacturer Narrative
Device lot number corrected from 19259204 to 19133634.Device expiration date corrected from 05/31/2019 to 04/30/2019.Device manufactured date corrected from 05/16/2016 to 04/12/2016.Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter in two pieces.There was contrast in the inflation lumen and blood in the guidewire lumen.The balloon was tightly folded.The shaft, hypotube, and bonds were microscopically and visually examined.There were numerous hypotube and shaft kinks throughout the device.There was also, a complete hypotube separation 55.5cm from the strain relief.The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The 90% stenosed, 18mm x 3.5mm target lesion was located in the severely tortuous and moderately calcified right coronary artery.A 4.0mm x 8mm quantum¿ maverick¿ balloon catheter was advanced to dilate the lesion.However, it was noted that the shaft had fractured.The device was completely removed from the patient's body and the procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key6178460
Report Number2134265-2016-11686
Device Sequence Number993536
Product Code LOX
UDI-Device Identifier08714729392606
UDI-Public(01)08714729392606(17)20190531(10)19259204
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2019
Device Model NumberH7493808008400
Device Catalogue Number38080-0840
Device Lot Number19133634
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/30/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/15/2016
Supplement Report FDA Received Date01/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-